RecallRadar

FDA Device recall Z-1222-2026

Brand Name: Single Use Electrosurgical Knife Product Name: Olympus Single Use Electrosurgical Knife KD-640L Model/Catalog Number: KD640-L P…

On 2026-02-11 the FDA classified a Class II recall of Brand Name: Single Use Electrosurgical Knife Product Name: Olympus Single Use Electrosurgical Knife KD-640L Model/Catalog Number: KD640-L P… by Olympus, citing A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1222-2026
ClassificationClass II
Statusongoing
Initiated2024-02-17
Published2026-02-11
Last updated2026-09-13 11:58 UTC
CompanyOlympus
DistributionUS

Product

Brand Name: Single Use Electrosurgical Knife Product Name: Olympus Single Use Electrosurgical Knife KD-640L Model/Catalog Number: KD640-L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.

Reason

A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.

Identifiers

code_info
Model/Catalog Number: KD640-L UDI: 04953170208423 Lot Number(s): 2ZK, 31K, 32K, 33K, 34K, 35K, 36K, 37K, 38K, 39K,…Model/Catalog Number: KD640-L UDI: 04953170208423 Lot Number(s): 2ZK, 31K, 32K, 33K, 34K, 35K, 36K, 37K, 38K, 39K, 3XK, 3YK, 3ZK, 41K, 42K, 43K, 44K, 45K, 46K, 47K, 48K, 49K, 4XK, 4YK, 4ZK, 51K, 52K, 53K, 54K, 55K, 58K, 59K

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1222-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.