FDA Device recall Z-1222-2022
DYNEX Agility, Agility Analyzer, Model No. 67000
- FDA Device
- Class II
- ongoing
- Published 2022-06-08
On 2022-06-08 the FDA classified a Class II recall of DYNEX Agility, Agility Analyzer, Model No. 67000 by Dynex Technologies, citing control samples aspirated from wrong SmartKit on the Agility. Agility software was updated to v1.4.7 to resolve the issue. This leads to a risk that a control from another assay's….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1222-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2020-07-05 |
| Published | 2022-06-08 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Dynex Technologies |
| Distribution | US |
Product
DYNEX Agility, Agility Analyzer, Model No. 67000
Reason
Control samples aspirated from wrong SmartKit on the Agility. Agility software was updated to v1.4.7 to resolve the issue. This leads to a risk that a control from another assay's SmartKit will be used instead of the correct control, which may lead to delayed patient results.
Identifiers
| code_info | UDI-DI: 05060456180058 Software version v1.4.3 Serial numbers: 1GXA0021 1GXA0022 1GXA0023 1GXA0024 1GXA0025 1GXA0026…UDI-DI: 05060456180058 Software version v1.4.3 Serial numbers: 1GXA0021 1GXA0022 1GXA0023 1GXA0024 1GXA0025 1GXA0026 1GXA0027 1GXA0028 1GXA0029 1GXA0030 1GXA0031 1GXA0032 1GXA0033 1GXA0034 1GXA0035 1GXA0036 1GXA0037 1GXA0038 1GXA0039 1GXA0040 1GXA0041 1GXA0042 1GXA0043 1GXA0044 1GXA0045 1GXA0046 1GXA0047 1GXA0048 1GXA0049 1GXA0050 1GXA0051 1GXA0052 1GXA0053 1GXA0054 1GXA0055 1GXA0056 1GXA0057 1GXA0058 1GXA0059 1GXA0060 1GXA0061 1GXA0062 1GXA0063 1GXA0064 1GXA0065 1GXA0066 1GXA0067 1GXA0068 1GXA0069 1GXA0070 1GXA0071 1GXA0072 1GXA0073 1GXA0074 1GXA0075 1GXA0076 1GXA0077 1GXA0078 1GXA0079 1GXA0080 1GXA0081 1GXA0082 1GXA0083 1GXA0084 1GXA0085 1GXA0086 1GXA0087 1GXA0088 1GXA0089 1GXA0090 1GXA0091 1GXA0092 1GXA0093 1GXA0094 1GXA0095 1GXA0096 1GXA0097 1GXA0098 1GXA0099 1GXA0100 1GXA0101 1GXA0102 1GXA0103 1GXA0104 1GXA0105 1GXA0106 1GXA0107 1GXA0108 1GXA0109 1GXA0110 1GXA0111 1GXA0112 1GXA0113 1GXA0114 1GXA0115 1GXA0116 1GXA0117 1GXA0118 1GXA0119 1GXA0120 1GXA0121 1GXA0122 1GXA0123 1GXA012 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1222-2022%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.