FDA Device recall Z-1222-2019

VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended f…

On 2019-05-01 the FDA classified a Class II recall of VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended f… by Ortho Clinical Diagnostics, citing one of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being report….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1222-2019
ClassificationClass II
Statusterminated
Initiated2019-03-11
Published2019-05-01
Last updated2026-09-13 11:55 UTC
CompanyOrtho Clinical Diagnostics
DistributionUS

Product

VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents, and VITROS Immunodiagnostic Products Reagents

Reason

One of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being reported without an error code to alert the user.

Identifiers

code_infoSystems Running Software Version 3.4 & 3.4.1

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.