FDA Device recall Z-1217-2017
The Enzyme II Calibrator is an in vitro diagnostic product for the Alanine Aminotransferase (ALTI) on the Dimension¿ clinical chemistry sys…
- FDA Device
- Class III
- terminated
- Published 2017-02-22
On 2017-02-22 the FDA classified a Class III recall of The Enzyme II Calibrator is an in vitro diagnostic product for the Alanine Aminotransferase (ALTI) on the Dimension¿ clinical chemistry sys… by Siemens Healthcare Diagnostics, citing complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Enzyme II calibrator. Internal investigation has confirmed that patient….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1217-2017 |
| Classification | Class III |
| Status | terminated |
| Initiated | 2016-05-05 |
| Published | 2017-02-22 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
The Enzyme II Calibrator is an in vitro diagnostic product for the Alanine Aminotransferase (ALTI) on the Dimension¿ clinical chemistry system. For prescription use only.
Reason
Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Enzyme II calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.
Identifiers
| code_info | Product Lot: 4MD047, 5CD052, 5FD017, 5GD050, 5HD010, 5JD069, 5LD051, 6AD054 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.