FDA Device recall Z-1216-2018
The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control of the navigation software and creation of a path…
- FDA Device
- Class II
- terminated
- Published 2018-04-04
On 2018-04-04 the FDA classified a Class II recall of The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control of the navigation software and creation of a path… by Stryker Instruments Div, citing stryker Instruments and Stryker Leibinger GmbH & Co. KG are voluntarily recalling certain serial numbers Thoracic Pedicle Feelers that are used in conjunction with the Stryker¿ Na….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1216-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-10-16 |
| Published | 2018-04-04 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Stryker Instruments Div |
| Distribution | n/a |
Product
The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control of the navigation software and creation of a pathway into the pedicle. The instrument has three types of buttons to remotely control the software: the UP, DOWN and SELECT buttons. The Pedicle Feeler is laser marked in 10 mm intervals to help indicate the depth in which the feeler tip has been inserted into the pedicle. Next to the navigational information, these depth indicators on the feeler tip are also helpful in determining the appropriate pedicle screw length.
Reason
Stryker Instruments and Stryker Leibinger GmbH & Co. KG are voluntarily recalling certain serial numbers Thoracic Pedicle Feelers that are used in conjunction with the Stryker¿ Navigation System because the settings on the machine were higher than intended. This change led to increased energy directed onto the tip of the Thoracic Pedicle Feeler which caused material degradation where the laser marks exist. There is the potential for the tip of the Thoracic Pedicle Feeler to break.
Identifiers
| code_info | Serial Number(s) Affected: 10897, 10898, 10899, 10900, 10901, 10902, 10903, 10904, 10905, 10906, 10907, 10908, 10909, 1…Serial Number(s) Affected: 10897, 10898, 10899, 10900, 10901, 10902, 10903, 10904, 10905, 10906, 10907, 10908, 10909, 10910, 10911, 10912, 10913, 10914, 10917, 10919, 10920, 10921, 10922, 10923, 10925, 10926, 10927, 10928. Product Code (UDI): (01)04546540503824 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Other recalls by Stryker Instruments Div
- Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters) Scanning Label (component part num…
FDA Device · Class II · 2022-07-27 · Stryker Instruments Div - SurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090
FDA Device · Class II · 2022-07-27 · Stryker Instruments Div - Triton Sponge System (Software), GAUS-3, Software Versions 4.6.5, 4.8.7, 4.9.6
FDA Device · Class III · 2022-07-27 · Stryker Instruments Div - Stryker Zyphr Disposable Cranial Perforator, Large 14/11 mm- is a sterile, single use cutting accessory intended for cutting an 11 mm diame…
FDA Device · Class II · 2020-11-18 · Stryker Instruments Div - Neptune E-SEP 165mm Blade Electrode, Catalog Number 0703-165-001 - Product Usage: The Neptune E-SEP is an Integrated Smoke Evacuation Penci…