FDA Device recall Z-1210-2013
Stryker Howmedica Osteonics Corp
- FDA Device
- Class II
- terminated
- Published 2013-05-08
On 2013-05-08 the FDA classified a Class II recall of Stryker Howmedica Osteonics Corp by Stryker Howmedica Osteonics, citing stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect. Th….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1210-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-03-28 |
| Published | 2013-05-08 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Stryker Howmedica Osteonics |
| Distribution | US |
Product
Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular fractures of the Distal Femur, Proximal Humerus, Proximal Tibia, and the Distal Tibia. Per the operative technique, " A 4.3 mm Drill (REF 702743) is then used to drill the core hole for the locking screw. Using a fluoroscopy, check the correct depth of the drill, and measure the length of the screw. The Drill Sleeve should now be removed, and the correct length 5.0 mm Locking Screw is inserted using the Screwdriver T20 and Screw Holding Sleeve."
Reason
Stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect. The incorrect scale could result in the surgeon selecting a screw that is 10mm longer than intended.
Identifiers
| code_info | Catalog number 703615 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1210-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.