FDA Device recall Z-1207-2020
FlexTip Plus Epidural Catheterization Kit, Product Code MP-17019-TKP
- FDA Device
- Class II
- terminated
- Published 2020-02-19
On 2020-02-19 the FDA classified a Class II recall of FlexTip Plus Epidural Catheterization Kit, Product Code MP-17019-TKP by Arrow International, citing there is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1207-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-01-15 |
| Published | 2020-02-19 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Arrow International |
| Distribution | AK, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WY |
Product
FlexTip Plus Epidural Catheterization Kit, Product Code MP-17019-TKP
Reason
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
Identifiers
| code_info | Lots: 23F19C0216 23F19F0037 23F19G0368 23F19J0089 23F19K0199 23F19M0018 23F19E0350 23F19F0269 23F19H0131 23F19J0208 23F19L0230 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.