FDA Device recall Z-1203-2026
Chest Drainage Units and Accessories: Product Code Description 1. 8888571562 Cardinal Health Sentinel Seal CDU 2. 8888571513 Cardinal Healt…
- FDA Device
- Class II
- ongoing
- Published 2026-02-04
On 2026-02-04 the FDA classified a Class II recall of Chest Drainage Units and Accessories: Product Code Description 1. 8888571562 Cardinal Health Sentinel Seal CDU 2. 8888571513 Cardinal Healt… by Cardinal Health 200, citing the firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1203-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-12-26 |
| Published | 2026-02-04 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Cardinal Health 200 |
| Distribution | US |
Product
Chest Drainage Units and Accessories: Product Code Description 1. 8888571562 Cardinal Health Sentinel Seal CDU 2. 8888571513 Cardinal Health Sentinel Seal CDU, Dual Drain 3. 8888571489 Cardinal Health Sentinel Seal CDU, with Easy Change Connector 4. 8888571370 Cardinal Health Altitude CDU, Dry Suction, with Easy Change Connector 5. 8888571299 Cardinal Health Aqua-Seal CDU, Wet Suction 6. 8888571406 Cardinal Health Aqua-Seal CDU, Wet Suction, Dual Drain 7. 8888571315 Cardinal Health Aqua-Seal CDU, Wet Suction, with Easy Change Connector 8. 8884713100 Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL 9. 1814713105 OUS¿Only - Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL (Bottle only) 10. 1180571570 OUS¿Only - Cardinal Health Thora-Seal CDU, Two Chamber System, 2600 mL 11. 8884713308 Cardinal Health Thora-Seal CDU, Three Chamber System, 2500 mL 12. 8884713900 Cardinal Health Thora-Seal Replacement Collection Chamber, 2500 mL
Reason
The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate tidaling/bubbling, which may confound the clinician's assessment of device functionality and lead to inadequate/inappropriate treatment/therapy, delay to treatment/therapy, and prolonged hospitalization.
Identifiers
| code_info | All units affected by IFU update. UDI-DI: 1. 8888571562 Cardinal Health Sentinel Seal CDU 50192253003091 (CS) 1019225…All units affected by IFU update. UDI-DI: 1. 8888571562 Cardinal Health Sentinel Seal CDU 50192253003091 (CS) 10192253003093 (EA) 2. 8888571513 Cardinal Health Sentinel Seal CDU, Dual Drain 50192253003084 (CS) 10192253003086 (EA) 3. 8888571489 Cardinal Health Sentinel Seal CDU, with Easy Change Connector 50192253003077 (CS) 10192253003079 (EA) 4. 8888571370 Cardinal Health Altitude CDU, Dry Suction, with Easy Change Connector 50192253003039 (CS) 10192253003031 (EA) 5. 8888571299 Cardinal Health Aqua-Seal CDU, Wet Suction 50192253003015 (CS) 10192253003017 (EA) 6. 8888571406 Cardinal Health Aqua-Seal CDU, Wet Suction, Dual Drain 50192253003060 (CS) 10192253003062 (EA) 7. 8888571315 Cardinal Health Aqua-Seal CDU, Wet Suction, with Easy Change Connector 50192253003022 (CS) 10192253003024 (EA) 8. 8884713100 Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL 50192253002643 (CS) 10192253002645 (EA) 9. 1814713105 OUS¿Only - Cardinal Health Thora-Seal Basic CDU, One Chamber, |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Other recalls by Cardinal Health 200
- Presource Packs containing Medtronic Microstream Components: Intubation Kit Catalog Numbers AN23366084 AN23366085 AN2336608H AN2336608J Int…
FDA Device · Class II · 2026-10-14 · Cardinal Health 200 - Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT,GU DAVINCI , Catalog Number: PGVMDVBWU
FDA Device · Class II · 2026-10-07 · Cardinal Health 200 - Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT, GYN LAPAROSCOPY NEW HAN , Catalog Number: PB23GLNLAC
FDA Device · Class II · 2026-10-07 · Cardinal Health 200 - Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT, LAP CHOLE , Catalog Number: PGCGLCBAC
FDA Device · Class II · 2026-10-07 · Cardinal Health 200 - Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT, LAVH , Catalog Number: PB63VHSTC
FDA Device · Class II · 2026-10-07 · Cardinal Health 200
All Cardinal Health 200 recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1203-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.