RecallRadar

FDA Device recall Z-1200-2020

EPIDURAL CATHETERIZATION KIT, Product Code IT-05400-DC

On 2020-02-19 the FDA classified a Class II recall of EPIDURAL CATHETERIZATION KIT, Product Code IT-05400-DC by Arrow International, citing there is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1200-2020
ClassificationClass II
Statusterminated
Initiated2020-01-15
Published2020-02-19
Last updated2026-09-13 11:55 UTC
CompanyArrow International
DistributionAK, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WY

Product

EPIDURAL CATHETERIZATION KIT, Product Code IT-05400-DC

Reason

There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.

Identifiers

code_infoLots: 71F19B2176 71F19C0468 71F19G1135 71F19L1545

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1200-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.