RecallRadar

FDA Device recall Z-1197-2025

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia

On 2025-02-26 the FDA classified a Class II recall of Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia by Augustine Temperature Management, citing there is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mat….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1197-2025
ClassificationClass II
Statusongoing
Initiated2024-12-16
Published2025-02-26
Last updated2026-09-13 11:57 UTC
CompanyAugustine Temperature Management
DistributionUS

Product

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia

Reason

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

Identifiers

code_infoUDI-DI UDI-DI 00855913001817, Serial Numbers: 1304292, 1304293, 1304296, 1304297

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1197-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.