RecallRadar

FDA Device recall Z-1196-2025

VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001

On 2025-03-05 the FDA classified a Class I recall of VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001 by Ventec Life Systems, citing multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate ti….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1196-2025
ClassificationClass I
Statusongoing
Initiated2025-02-03
Published2025-03-05
Last updated2026-09-13 11:57 UTC
CompanyVentec Life Systems
DistributionCA, MO, NY

Product

VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000, PRT-01185-002

Reason

Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less inspiratory volume, less inspiratory pressure, and less PEEP delivered.

Identifiers

code_info
REF/UDI-DI/Serial Numbers: PRT-00490-001/00855573007792/123627, 121892, 119639; PRT-01185-000/00855573007877/5038421,…REF/UDI-DI/Serial Numbers: PRT-00490-001/00855573007792/123627, 121892, 119639; PRT-01185-000/00855573007877/5038421, 116700, 117933, 5038112, 5038619, 5038717, 5038953, 5039086, 5039091, 5039361, 5039499, 5039503, 5039541, 5039636, 5039938, 5040019, 5040640, 5040762, 5041060, 5041078, 5041534, 5042558, 117344, 118464, 118509, 5039387; PRT-01185-002/00850018761154/5037655