FDA Device recall Z-1192-2015
NAMIC Convenience Kit, LEFT HEART KIT, UPN H965600125911, REF/Catalog No. 60012591, STERILE, Rx ONLY ---For single use only. --- Device Lis…
- FDA Device
- Class II
- terminated
- Published 2015-03-04
On 2015-03-04 the FDA classified a Class II recall of NAMIC Convenience Kit, LEFT HEART KIT, UPN H965600125911, REF/Catalog No. 60012591, STERILE, Rx ONLY ---For single use only. --- Device Lis… by Navilyst Medical, citing the labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1192-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-01-09 |
| Published | 2015-03-04 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Navilyst Medical |
| Distribution | US |
Product
NAMIC Convenience Kit, LEFT HEART KIT, UPN H965600125911, REF/Catalog No. 60012591, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
Reason
The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
Identifiers
| code_info | Lot 4440346 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1192-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.