RecallRadar

FDA Device recall Z-1185-2016

Transport Stretcher, and Head and Neck Surgery Stretcher Non-powered stretcher

On 2016-03-23 the FDA classified a Class II recall of Transport Stretcher, and Head and Neck Surgery Stretcher Non-powered stretcher by Stryker Medical Division, citing the hydraulic jack assemblies (PN: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) were manufactured using the incorrect oil. The oil used does not contain the same we….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1185-2016
ClassificationClass II
Statusterminated
Initiated2016-02-05
Published2016-03-23
Last updated2026-09-13 11:54 UTC
CompanyStryker Medical Division
DistributionAL, CA, DC, FL, MN, MT, NC, NE, NY, PA, RI, TN, TX, UT, WV

Product

Transport Stretcher, and Head and Neck Surgery Stretcher Non-powered stretcher

Reason

The hydraulic jack assemblies (PN: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) were manufactured using the incorrect oil. The oil used does not contain the same wear resistance additives as the specified oils and therefore its use increases the potential for wear which may cause a leak event or jack drift. Testing was done to evaluate the jacks with incorrect oil and leak events we

Identifiers

code_info
The hydraulic jack assemblies (Part numbers: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) used as servic…The hydraulic jack assemblies (Part numbers: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) used as service parts on Stretchers assembled from 8/28/15 - 8/29/15

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1185-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.