FDA Device recall Z-1184-2013
Medi-Trace Cadence Adult Radiotransparent Defibrillation Electrode Product ID: 22550R. Intended for use in cardioversion, pacing, ECG monit…
- FDA Device
- Class II
- terminated
- Published 2013-05-08
On 2013-05-08 the FDA classified a Class II recall of Medi-Trace Cadence Adult Radiotransparent Defibrillation Electrode Product ID: 22550R. Intended for use in cardioversion, pacing, ECG monit… by Covidien, citing defibrillation Electrodes may arc/spark result in thermal damage to the leadwire, which could render the device incapable of delivering the appropriate energy or shock to the pati….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1184-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-04-08 |
| Published | 2013-05-08 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Covidien |
| Distribution | US |
Product
Medi-Trace Cadence Adult Radiotransparent Defibrillation Electrode Product ID: 22550R. Intended for use in cardioversion, pacing, ECG monitoring, and defibrillation procedures.
Reason
Defibrillation Electrodes may arc/spark result in thermal damage to the leadwire, which could render the device incapable of delivering the appropriate energy or shock to the patient
Identifiers
| code_info | Lot Numbers: 226543X, 230054X , 230814X, 232182X, 232805X, 234245X, 304667X, 301844X, 300446X, 305320X |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1184-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.