Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (ca…

On 2024-02-28 the FDA classified a Class II recall of Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (ca… by Abiomed, citing retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1175-2024
ClassificationClass II
Statusterminated
Initiated2021-03-01
Published2024-02-28
Last updated2026-09-13 11:57 UTC
CompanyAbiomed
DistributionUS

Product

Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular)

Reason

Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.

Identifiers

code_info
Product Code 0042-0000-US UDI-DI 00813502010022; Product Code 0042-0010-US UDI-DI 00813502010985; Product Code 0042-004…Product Code 0042-0000-US UDI-DI 00813502010022; Product Code 0042-0010-US UDI-DI 00813502010985; Product Code 0042-0040-US UDI-DI 00813502011401; Software Versions V8.4 and V8.4.1

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1175-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.