Toshiba DRAD-3000E FPD Wireless System Product Usage The DRAD-3000¿ is intended for use with the ceiling-suspended tube support, high volta…

On 2016-03-30 the FDA classified a Class II recall of Toshiba DRAD-3000E FPD Wireless System Product Usage The DRAD-3000¿ is intended for use with the ceiling-suspended tube support, high volta… by Toshiba American Medical Systems, citing when a user performed radiography using the wireless FPD, a message window appeared on the monitor stating that imaging transmission was not completed with no radiographic image.….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1175-2016
ClassificationClass II
Statusterminated
Initiated2016-01-14
Published2016-03-30
Last updated2026-09-13 11:54 UTC
CompanyToshiba American Medical Systems
DistributionUS

Product

Toshiba DRAD-3000E FPD Wireless System Product Usage The DRAD-3000¿ is intended for use with the ceiling-suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck, and limbs. This system is used for image acquisition, image display and transmission/output or images to external devices.

Reason

When a user performed radiography using the wireless FPD, a message window appeared on the monitor stating that imaging transmission was not completed with no radiographic image. It also indicated to select the "OK" button to re-acquire the image data or to select the¿'Cancel" button to cancel the re-acquisition. As instructed the user selected "OK" and the same message window appeared. This same operation was repeated several times with the identical result - no image. Finally, the user selected "Cancel" and the re-acquisition was terminated.

Identifiers

code_infoSerial number : USB1522023

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1175-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.