FDA Device recall Z-1171-2024

Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images

On 2024-02-28 the FDA classified a Class II recall of Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images by Fujifilm Healthcare Americas, citing the LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1171-2024
ClassificationClass II
Statusongoing
Initiated2024-01-03
Published2024-02-28
Last updated2026-09-13 11:57 UTC
CompanyFujifilm Healthcare Americas
DistributionUS

Product

Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images.

Reason

The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.

Identifiers

code_infoUDI-DI 854904006008 Software versions 7.3.000, 7.2.200, 7.2.100, 7.2.000, 7.1.000, 7.0.000