FDA Device recall Z-1169-2025

Vios Monitoring System Bedside Monitor, Model BSM2050, Software Version Numbers 1.54.20 and 1.56.00

On 2025-02-19 the FDA classified a Class II recall of Vios Monitoring System Bedside Monitor, Model BSM2050, Software Version Numbers 1.54.20 and 1.56.00 by Murata Vios, citing A premature failure condition was identified with the L2050R pulse oximeter SpO2 sensor that connects to the L2050G Vios chest sensor device. When a failure occurs, a Pulse Ox Se….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1169-2025
ClassificationClass II
Statusongoing
Initiated2023-06-06
Published2025-02-19
Last updated2026-09-13 11:57 UTC
CompanyMurata Vios
DistributionUS

Product

Vios Monitoring System Bedside Monitor, Model BSM2050, Software Version Numbers 1.54.20 and 1.56.00

Reason

A premature failure condition was identified with the L2050R pulse oximeter SpO2 sensor that connects to the L2050G Vios chest sensor device. When a failure occurs, a Pulse Ox Sensor Off warning condition will be incorrectly displayed by the system.

Identifiers

code_infoAll systems with Software Version Numbers 1.54.20 and 1.56.00 affected; UDI-DI: B595BSM20500

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.