FDA Device recall Z-1164-2025
Microbiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet), REF 8235
- FDA Device
- Class II
- ongoing
- Published 2025-02-19
On 2025-02-19 the FDA classified a Class II recall of Microbiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet), REF 8235 by Microbiologics, citing internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have faile….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1164-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-01-20 |
| Published | 2025-02-19 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Microbiologics |
| Distribution | US |
Product
Microbiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet), REF 8235; in-vitro diagnostic test
Reason
Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.
Identifiers
| code_info | UDI/DI 70845357043077, Lot Numbers: 8235-32, 8235-33 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1164-2025%22