RecallRadar

FDA Device recall Z-1163-2025

Microbiologics Multi-Strain Human Papillomavirus (HPV) Control Panel (Inactivated Pellet), REF 8234

On 2025-02-19 the FDA classified a Class II recall of Microbiologics Multi-Strain Human Papillomavirus (HPV) Control Panel (Inactivated Pellet), REF 8234 by Microbiologics, citing internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have faile….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1163-2025
ClassificationClass II
Statusongoing
Initiated2025-01-20
Published2025-02-19
Last updated2026-09-13 11:57 UTC
CompanyMicrobiologics
DistributionUS

Product

Microbiologics Multi-Strain Human Papillomavirus (HPV) Control Panel (Inactivated Pellet), REF 8234; in-vitro diagnostic test

Reason

Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.

Identifiers

code_infoUDI/DI 70845357043077, Lot Number 8234-22

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1163-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.