FDA Device recall Z-1160-2013
GemStar Pump - Battery Product Usage: The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the h…
- FDA Device
- Class I
- terminated
- Published 2013-05-08
On 2013-05-08 the FDA classified a Class I recall of GemStar Pump - Battery Product Usage: The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the h… by Hospira, citing battery level lower then 2.4 volts results in corrupt history log and loss of data for preview therapy's parameters and causes an 11/004 error resulting in the pump not being able….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1160-2013 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2013-03-18 |
| Published | 2013-05-08 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Hospira |
| Distribution | US |
Product
GemStar Pump - Battery Product Usage: The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline batteries. When powered by batteries, The GemStar Pump is ideal for ambulatory patients.
Reason
Battery level lower then 2.4 volts results in corrupt history log and loss of data for preview therapy's parameters and causes an 11/004 error resulting in the pump not being able to be used.
Identifiers
| code_info | List Numbers: 13086, 13087, 13088 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1160-2013%22