RecallRadar

FDA Device recall Z-1158-2014

Smith & Nephew RENASYS-F Foam Dressing Kit with Soft Port REF #'s: 66800794, 66800795, and 66800796 (The RENASYS Soft Port Dressing Kits ar…

On 2014-03-19 the FDA classified a Class II recall of Smith & Nephew RENASYS-F Foam Dressing Kit with Soft Port REF #'s: 66800794, 66800795, and 66800796 (The RENASYS Soft Port Dressing Kits ar… by Smith And Nephew, citing reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1158-2014
ClassificationClass II
Statusterminated
Initiated2014-01-09
Published2014-03-19
Last updated2026-09-13 11:53 UTC
CompanySmith And Nephew
DistributionUS

Product

Smith & Nephew RENASYS-F Foam Dressing Kit with Soft Port REF #'s: 66800794, 66800795, and 66800796 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)

Reason

Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo

Identifiers

code_infoALL LOT #'s: 2013011828 and lower- Small, 2013010125 and lower- Medium, and 2013011692 and lower - Large

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1158-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.