FDA Device recall Z-1148-2020
Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Single, REF 3910-500-471, UDI 07613327464818, Sterile R, RX Only CE 0197
- FDA Device
- Class II
- terminated
- Published 2020-02-19
On 2020-02-19 the FDA classified a Class II recall of Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Single, REF 3910-500-471, UDI 07613327464818, Sterile R, RX Only CE 0197 by Stryker, citing complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1148-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-07-19 |
| Published | 2020-02-19 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Stryker |
| Distribution | AK, AR, AZ, CA, CO, FL, GA, ID, IL, KY, MD, ME, NJ, NV, NY, OH, TX, UT, WI |
Product
Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Single, REF 3910-500-471, UDI 07613327464818, Sterile R, RX Only CE 0197
Reason
Complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver. This could result in an additional anchor insertion site in the bone being used.
Identifiers
| code_info | 19113AG2, 19130AG2, 19134AG2, 19115AG2, 19131AG2, 19133AG2, 19154AG2,19155AG2, 19156AG2, 19157AG2, 19149AG2. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1148-2020%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.