FDA Device recall Z-1147-2019
Tina-quant IgM Gen.2, Catalog Nos. 03507190190, 05220726190, 03507041190, 03507149190 Product Usage: Immunoturbidimetric assay for the quan…
- FDA Device
- Class II
- terminated
- Published 2019-04-24
On 2019-04-24 the FDA classified a Class II recall of Tina-quant IgM Gen.2, Catalog Nos. 03507190190, 05220726190, 03507041190, 03507149190 Product Usage: Immunoturbidimetric assay for the quan… by Roche Diagnostics, citing new endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D mo….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1147-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-12-17 |
| Published | 2019-04-24 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Roche Diagnostics |
| Distribution | US |
Product
Tina-quant IgM Gen.2, Catalog Nos. 03507190190, 05220726190, 03507041190, 03507149190 Product Usage: Immunoturbidimetric assay for the quantitative in vitro determination of IgM in human serum and plasma on automated clinical chemistry analyzers analyzers.
Reason
New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS INTEGRA 400 plus analyzers are also impacted for BILT3 and LACT2 assays***