RecallRadar

FDA Device recall Z-1143-2023

ADVIA Centaur XP and ADVIA Centaur XPT Anti-HBe2 (aHBe2) 50T- In vitro diagnostic immunoassay for the qualitative detection of antibodies t…

On 2023-03-01 the FDA classified a Class II recall of ADVIA Centaur XP and ADVIA Centaur XPT Anti-HBe2 (aHBe2) 50T- In vitro diagnostic immunoassay for the qualitative detection of antibodies t… by Siemens Healthcare Diagnostics, citing anti-HBe2 (aHBe2) interference with the Hepatitis B e Antigen (HBeAg) Assay, leads to elevated or reactive Hepatitis B e Antigen (HBeAg) results on samples that are non-reactive w….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1143-2023
ClassificationClass II
Statusongoing
Initiated2022-12-22
Published2023-03-01
Last updated2026-09-13 11:56 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

ADVIA Centaur XP and ADVIA Centaur XPT Anti-HBe2 (aHBe2) 50T- In vitro diagnostic immunoassay for the qualitative detection of antibodies to the e antigen of the hepatitis B virus (HBV) in human pediatric (2-21 years old) and adult serum, EDTA plasma, or lithium heparin. Siemens Material Number (SMN): 10720831

Reason

Anti-HBe2 (aHBe2) interference with the Hepatitis B e Antigen (HBeAg) Assay, leads to elevated or reactive Hepatitis B e Antigen (HBeAg) results on samples that are non-reactive when HBeAg testing follows aHBe2 testing, regardless of whether both tests are performed on the same sample

Identifiers

code_info
UDI: (01)00630414167923(10)B34847(17)20230309 (01)00630414167923(10)B34848(17)20230309 (01)00630414167923(10)B35113(17)…UDI: (01)00630414167923(10)B34847(17)20230309 (01)00630414167923(10)B34848(17)20230309 (01)00630414167923(10)B35113(17)20230629. Lot Numbers: B34847, B34848, B35113. Affects all current and future lots of the ADVIA Centaur XP and ADVIA Centaur XPT HBeAg assay until a solution is Implemented.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1143-2023%22