FDA Device recall Z-1143-2015

Starburst MRI Semi-Flex Electrode Device, 25cm/Attached Cable/14Ga/6.4F, Product No. H7877001039091, Catalog No. 700-103909

On 2015-03-04 the FDA classified a Class II recall of Starburst MRI Semi-Flex Electrode Device, 25cm/Attached Cable/14Ga/6.4F, Product No. H7877001039091, Catalog No. 700-103909 by Angiodynamics, citing printed mark on the trocar intended to indicate a 5cm distance from the distal tip of the needle is actually located 4cm from the distal tip of the needle.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1143-2015
ClassificationClass II
Statusterminated
Initiated2014-12-19
Published2015-03-04
Last updated2026-09-13 11:53 UTC
CompanyAngiodynamics
DistributionAR, AZ, FL, MI, MN, MO, NY, TX, VA, WA, WI

Product

Starburst MRI Semi-Flex Electrode Device, 25cm/Attached Cable/14Ga/6.4F, Product No. H7877001039091, Catalog No. 700-103909; 25cm/14Ga/6.4F, Product No. H7877001026151, Catalog No. 700-102615; and 25cm/14Ga/6.4F, Product No. H7877001026151, Catalog No. 700-102615

Reason

Printed mark on the trocar intended to indicate a 5cm distance from the distal tip of the needle is actually located 4cm from the distal tip of the needle.

Identifiers

code_infoLot No. 4773676, 4776428 and 4780110