FDA Device recall Z-1142-2024
IonoStar Plus, Catalog (REF) Numbers: a) 2540, b)2543, c) 2544, d) 2545, e) 2546, f) 2547, g) 2549
- FDA Device
- Class II
- ongoing
- Published 2024-02-28
On 2024-02-28 the FDA classified a Class II recall of IonoStar Plus, Catalog (REF) Numbers: a) 2540, b)2543, c) 2544, d) 2545, e) 2546, f) 2547, g) 2549 by Voco, citing due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive compone….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1142-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-01-25 |
| Published | 2024-02-28 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Voco |
| Distribution | IA, NY, PA, WA |
Product
IonoStar Plus, Catalog (REF) Numbers: a) 2540, b)2543, c) 2544, d) 2545, e) 2546, f) 2547, g) 2549; dental cement
Reason
Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the instructions for use is not possible.
Identifiers
| code_info | a) 2540, UDI/DI +E22125401, Lot Numbers: 2338699, 2347154; b)2543, UDI/DI +E22125431, Lot Numbers: 2344046; c) 2544…a) 2540, UDI/DI +E22125401, Lot Numbers: 2338699, 2347154; b)2543, UDI/DI +E22125431, Lot Numbers: 2344046; c) 2544, UDI/DI +E22125441, Lot Numbers: 2342759, 2439537; d) 2545, UDI/DI +E22125451, Lot Numbers: 2348503; e) 2546, UDI/DI +E22125461, Lot Numbers: 2344045, 2349452; f) 2547, UDI/DI +E22125471, Lot Numbers: 2349453; g) 2549, UDI/DI +E22125491, Lot Numbers: 2342760, 2344016, 2344047, 2351233 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1142-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.