FDA Device recall Z-1141-2017

Dimension Vista¿ Ammonia Flex¿ reagent cartridge / AMM Dimension: The AMM method is an in vitro diagnostic test for the quantitative measur…

On 2017-02-15 the FDA classified a Class II recall of Dimension Vista¿ Ammonia Flex¿ reagent cartridge / AMM Dimension: The AMM method is an in vitro diagnostic test for the quantitative measur… by Siemens Healthcare Diagnostics, citing siemens Healthcare Diagnostics has determined that Dimension Vista AMM (K3119) Flex reagent cartridge flex lots 16187BE, 16225BB, 16265AB do not meet the 60-day calibration interv….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1141-2017
ClassificationClass II
Statusterminated
Initiated2016-12-16
Published2017-02-15
Last updated2026-09-13 11:54 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Dimension Vista¿ Ammonia Flex¿ reagent cartridge / AMM Dimension: The AMM method is an in vitro diagnostic test for the quantitative measurement of ammonia in human plasma on the Dimension clinical chemistry system. Ammonia measurements are used in the diagnosis and treatment of severe liver disorders such as cirrhosis, hepatitis and Reye s syndrome

Reason

Siemens Healthcare Diagnostics has determined that Dimension Vista AMM (K3119) Flex reagent cartridge flex lots 16187BE, 16225BB, 16265AB do not meet the 60-day calibration interval claim due to reagent instability and results may show an Abnormal Assay.

Identifiers

code_infolots 16187BE, 16225BB and 16265AB

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.