FDA Device recall Z-1140-2019
Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector
- FDA Device
- Class II
- terminated
- Published 2019-04-24
On 2019-04-24 the FDA classified a Class II recall of Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector by Philips Medical Systems Nederlands, citing additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch peda….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1140-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-02-19 |
| Published | 2019-04-24 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Philips Medical Systems Nederlands |
| Distribution | US |
Product
Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 Product The Allura Xper FD Series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures.
Reason
Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.
Identifiers
| code_info | Foot Switches: 452270000141 Footswitch CV 3p 4m; 452270000151 Footswitch MD 3p 6m; 452270000142 Footswitch CV 3p 4m;…Foot Switches: 452270000141 Footswitch CV 3p 4m; 452270000151 Footswitch MD 3p 6m; 452270000142 Footswitch CV 3p 4m; 452270000381 Footswitch CV 3p 8m; 452270000382 Footswitch CV 3p 8m; 459800076001 Footswitch CV 4p+2 4m; 459800076021 Footswitch CV 4p+2 8m; 459800415571 Wireless Footswitch 4p+2; 459800415581 Wireless footswitch set 4p+2; 459800415531 Wireless Footswitch 3P; 459800415561 Wireless footswitch set 3P. |
|---|