RecallRadar

FDA Device recall Z-1138-2023

NovaStar TS, Standard elbow, Reusable, Material Numbers MP01579 (Size S), MP01580 (Size M), and MP01581 (Size L)

On 2023-03-01 the FDA classified a Class I recall of NovaStar TS, Standard elbow, Reusable, Material Numbers MP01579 (Size S), MP01580 (Size M), and MP01581 (Size L) by Draeger Medical, citing the magnets used in the swivel frame/tube set and the headgear clips might impact medical devices (implanted as well as not implanted) and metallic implants.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1138-2023
ClassificationClass I
Statusongoing
Initiated2023-01-23
Published2023-03-01
Last updated2026-09-13 11:56 UTC
CompanyDraeger Medical
DistributionUS

Product

NovaStar TS, Standard elbow, Reusable, Material Numbers MP01579 (Size S), MP01580 (Size M), and MP01581 (Size L)

Reason

The magnets used in the swivel frame/tube set and the headgear clips might impact medical devices (implanted as well as not implanted) and metallic implants.

Identifiers

code_infoAll Lot numbers. UDI-DI: 04048675040552 (MP01579); 04048675040569 (MP01580); 04048675040538 (MP01581)

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.