RecallRadar

FDA Device recall Z-1138-2019

IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and managemen…

On 2019-04-24 the FDA classified a Class II recall of IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and managemen… by Topcon Medical Systems, citing when a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1138-2019
ClassificationClass II
Statusterminated
Initiated2019-03-06
Published2019-04-24
Last updated2026-09-13 11:55 UTC
CompanyTopcon Medical Systems
DistributionUS

Product

IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices without controlling or altering the functions and parameters of any medical devices or through computerized networks.

Reason

When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information such as name, patient's ID, etc. on the monitor screen of the PC used.

Identifiers

code_infoSystem Version 1.53

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1138-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.