FDA Device recall Z-1138-2016
Product 40 consists of all product under product code: HSB and same usage: Item no: 225706565 ITST ANTI-ROTATION SCREW 225707065 ITST ANTI-…
- FDA Device
- Class II
- terminated
- Published 2016-03-23
On 2016-03-23 the FDA classified a Class II recall of Product 40 consists of all product under product code: HSB and same usage: Item no: 225706565 ITST ANTI-ROTATION SCREW 225707065 ITST ANTI-… by Zimmer Manufacturing B V, citing as a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1138-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-01-11 |
| Published | 2016-03-23 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Zimmer Manufacturing B V |
| Distribution | US |
Product
Product 40 consists of all product under product code: HSB and same usage: Item no: 225706565 ITST ANTI-ROTATION SCREW 225707065 ITST ANTI-ROTATION SCREW 225708065 ITST ANTI-ROTATION SCREW 225708565 ITST ANTI-ROTATION SCREW 225709065 ITST ANTI-ROTATION SCREW Product Usage: An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Reason
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Identifiers
| code_info | lot no.: 62208147 62208146 62208148 62042715 62042716 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1138-2016%22