RecallRadar

FDA Device recall Z-1136-2026

TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00…

On 2026-01-28 the FDA classified a Class II recall of TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00… by Sysmex America, citing under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1136-2026
ClassificationClass II
Statusongoing
Initiated2025-12-15
Published2026-01-28
Last updated2026-09-13 11:58 UTC
CompanySysmex America
DistributionUS

Product

TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00-17

Reason

Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.

Identifiers

code_info
Serial Numbers: 11281 11611 11918 11790 12040 11991 12218 11950 11795 11533 11635 12054 11610 11453 11753 11539 11990 1…Serial Numbers: 11281 11611 11918 11790 12040 11991 12218 11950 11795 11533 11635 12054 11610 11453 11753 11539 11990 12298 11774 11541 11981 11788 11834 12005 11907 11524 12237 11246 12154 12104 12299 12176 11711 11420 12153 12177 12074 11405 11303 12039 11250 11336 12137 12075

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1136-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.