RecallRadar

FDA Device recall Z-1136-2021

VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWe…

On 2021-03-03 the FDA classified a Class II recall of VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWe… by Ortho Clinical Diagnostics, citing A deformity affecting a specific lot of VITROS¿ Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1136-2021
ClassificationClass II
Statusterminated
Initiated2021-01-14
Published2021-03-03
Last updated2026-09-13 11:56 UTC
CompanyOrtho Clinical Diagnostics
DistributionUS

Product

VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In addition, VersaTips are used to dispense R1 Reagent Fluid into cuvettes in combination with sample fluid for Micro Tips Assays.

Reason

A deformity affecting a specific lot of VITROS¿ Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays.

Identifiers

code_infoLot 413420 Model(s): VITROS 250, VITROS 350, VITROS ECi, VITROS 5,1 FS, VITROS 3600, VITROS 4600, VITROS 5600, VITROS XT7600.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1136-2021%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.