FDA Device recall Z-1134-2018
Medfusion Syringe Pump 4000 Series The Medfusion¿ Syringe Infusion Pumps are indicated for the following uses: 1) Administration of fluids…
- FDA Device
- Class II
- terminated
- Published 2018-03-28
On 2018-03-28 the FDA classified a Class II recall of Medfusion Syringe Pump 4000 Series The Medfusion¿ Syringe Infusion Pumps are indicated for the following uses: 1) Administration of fluids… by Smiths Medical Asd, citing certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1134-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-11-13 |
| Published | 2018-03-28 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Smiths Medical Asd |
| Distribution | US |
Product
Medfusion Syringe Pump 4000 Series The Medfusion¿ Syringe Infusion Pumps are indicated for the following uses: 1) Administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids; 2) By the following delivery routes: arterial, epidural, intravenous, intrathecal, subcutaneous, and enteral; 3)By the following delivery modes: continuous, volume/time, mass, body weight, intermittent, and bolus; 4) In critical care, anesthesia, neonatal, and pediatric applications or other healthcare settings where use of the syringe infusion pump can be monitored or supervised by a clinician; 5) Inside the MRI room mounted outside the 150 Gauss line and with shielded magnets of field strength of 1.5 Tesla.
Reason
Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump) results in an inability to complete pump programming. Misidentification of a syringe is where the pump misinterprets the syringe size.
Identifiers
| code_info | Product Codes: 4000-0101-50, 4000-0101-51, 4000-0105-50, 4000-0105-51, 4000-0150-78, 4000-0106-00, 4000-0106-01 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1134-2018%22