RecallRadar

FDA Device recall Z-1134-2014

DeRoyal ST. Circumcision Clamp, REF 32-1622, Size 1.3 cm, 1 Per Pack, Rx Only STERILE EO, distributed by DeRoyal Industries

On 2014-03-12 the FDA classified a Class II recall of DeRoyal ST. Circumcision Clamp, REF 32-1622, Size 1.3 cm, 1 Per Pack, Rx Only STERILE EO, distributed by DeRoyal Industries by Deroyal Industries, citing three lots of circumcision clamps were manufactured out of specification and as a result may not properly clamp at the surgical site completely.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1134-2014
ClassificationClass II
Statusterminated
Initiated2014-01-16
Published2014-03-12
Last updated2026-09-13 11:53 UTC
CompanyDeroyal Industries
DistributionAL, FL, IL, KS, NJ, OK, OR, TX, VA, WA

Product

DeRoyal ST. Circumcision Clamp, REF 32-1622, Size 1.3 cm, 1 Per Pack, Rx Only STERILE EO, distributed by DeRoyal Industries.

Reason

Three lots of circumcision clamps were manufactured out of specification and as a result may not properly clamp at the surgical site completely.

Identifiers

code_infolot numbers 33347799, 33444071, and 33423747

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.