SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters

On 2013-04-24 the FDA classified a Class II recall of SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters by Sr Instruments, citing SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile t….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1128-2013
ClassificationClass II
Statusterminated
Initiated2012-12-13
Published2013-04-24
Last updated2026-09-13 11:53 UTC
CompanySr Instruments
DistributionUS

Product

SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# FG3105 Patient Scales. Product Usage: The FG3105 Scale is designed for Joerns Healthcare for use in portable patient weighing systems. The scale is designed to be connected between the patient hoist mechanism and patient sling apparatus. The scale is not intended to be used for patient transport.

Reason

SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile type load cell involving the FG3105 Patient Scale. The failures resulted in patient falls.

Identifiers

code_infoSerial numbers: 1005 thru 1780, 1785 thru 1809, 1866 thru 1868 and 1870 thru 1875.