FDA Device recall Z-1128-2013
SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters
- FDA Device
- Class II
- terminated
- Published 2013-04-24
On 2013-04-24 the FDA classified a Class II recall of SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters by Sr Instruments, citing SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile t….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1128-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-12-13 |
| Published | 2013-04-24 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Sr Instruments |
| Distribution | US |
Product
SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# FG3105 Patient Scales. Product Usage: The FG3105 Scale is designed for Joerns Healthcare for use in portable patient weighing systems. The scale is designed to be connected between the patient hoist mechanism and patient sling apparatus. The scale is not intended to be used for patient transport.
Reason
SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile type load cell involving the FG3105 Patient Scale. The failures resulted in patient falls.
Identifiers
| code_info | Serial numbers: 1005 thru 1780, 1785 thru 1809, 1866 thru 1868 and 1870 thru 1875. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1128-2013%22