FDA Device recall Z-1126-2018
CATHETER REPAIR KIT, REF RMS52001, UDI 884908039658 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female…
- FDA Device
- Class II
- terminated
- Published 2018-03-28
On 2018-03-28 the FDA classified a Class II recall of CATHETER REPAIR KIT, REF RMS52001, UDI 884908039658 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female… by Medical Components, citing there is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1126-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-02-23 |
| Published | 2018-03-28 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Medical Components |
| Distribution | US |
Product
CATHETER REPAIR KIT, REF RMS52001, UDI 884908039658 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Reason
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Identifiers
| code_info | Lot Numbers: MCBX400, MCFA150, MCFT460, MCCHOlO, MCDN490, MCCV190, MCCZ680, MCDDlOO, MCDV170, MCDW310 |
|---|