FDA Device recall Z-1126-2013

Accuray CyberKnife System Medical Physicist, Robotic Radiosurgery System, Medical charged-particle radiation therapy system, Model: Iris Va…

On 2013-04-24 the FDA classified a Class II recall of Accuray CyberKnife System Medical Physicist, Robotic Radiosurgery System, Medical charged-particle radiation therapy system, Model: Iris Va… by Accuray, citing A defect on the master-side connector block of the pneumatic tool changer could introduce a tilt in the collimator housings, resulting in a potential shift to the central axis of….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1126-2013
ClassificationClass II
Statusterminated
Initiated2013-03-19
Published2013-04-24
Last updated2026-09-13 11:53 UTC
CompanyAccuray
DistributionUS

Product

Accuray CyberKnife System Medical Physicist, Robotic Radiosurgery System, Medical charged-particle radiation therapy system, Model: Iris Variable Aperture Collimator P/Ns 028986 and 031777. The CyberKnife System is indicated for radiation treatment.

Reason

A defect on the master-side connector block of the pneumatic tool changer could introduce a tilt in the collimator housings, resulting in a potential shift to the central axis of the radiation beam. The issue could pose significant clinical impact to plans created using the smallest collimators.

Identifiers

code_infoAll units of these models.