FDA Device recall Z-1125-2018

TWO PART ADAPTER, REF MCTLA2000, UDI 884908027310 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female he…

On 2018-03-28 the FDA classified a Class II recall of TWO PART ADAPTER, REF MCTLA2000, UDI 884908027310 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female he… by Medical Components, citing there is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1125-2018
ClassificationClass II
Statusterminated
Initiated2018-02-23
Published2018-03-28
Last updated2026-09-13 11:54 UTC
CompanyMedical Components
DistributionUS

Product

TWO PART ADAPTER, REF MCTLA2000, UDI 884908027310 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.

Reason

There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.

Identifiers

code_infoLot Numbers: MCFF700, MCFB160, MKAG11O, MCDP590, MCDS140 & MCFSOSO