FDA Device recall Z-1119-2026
ICU Medical Plum Duo Precision IV Pump, List Numbers: 400020403 (US), 400021003 (Philippines)
- FDA Device
- Class II
- ongoing
- Published 2026-01-28
On 2026-01-28 the FDA classified a Class II recall of ICU Medical Plum Duo Precision IV Pump, List Numbers: 400020403 (US), 400021003 (Philippines) by Icu Medical, citing plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is deliver….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1119-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-12-15 |
| Published | 2026-01-28 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Icu Medical |
| Distribution | US |
Product
ICU Medical Plum Duo Precision IV Pump, List Numbers: 400020403 (US), 400021003 (Philippines)
Reason
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.
Identifiers
| code_info | 400020403 (US), Software Versions: 1.2.2; 400021003 (Philippines), Software Versions: 1.2.0 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1119-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.