FDA Device recall Z-1119-2017
IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LKUP1, SMN 10380956 (OUS Kit Catalog) Product Usage: For in vitro…
- FDA Device
- Class II
- terminated
- Published 2017-02-08
On 2017-02-08 the FDA classified a Class II recall of IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LKUP1, SMN 10380956 (OUS Kit Catalog) Product Usage: For in vitro… by Siemens Healthcare Diagnostics, citing negative bias with IMMULITE¿ /IMMULITE¿ 1000 Third Generation PSA (prostate-specific antigen) kit. The investigation by Siemens has confirmed that patient samples exhibit an avera….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1119-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-12-21 |
| Published | 2017-02-08 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LKUP1, SMN 10380956 (OUS Kit Catalog) Product Usage: For in vitro diagnostic use with the IMMULITE¿ and IMMULITE 1000 Analyzers for the quantitative measurement of prostate-specific antigen (PSA) in human serum, as an aid in the detection of prostate cancer when used in conjunction with digital rectal examination (DRE) in men aged 50 years or older. This assay is further indicated as an adjunctive test to aid in the management of prostate cancer patients.
Reason
Negative bias with IMMULITE¿ /IMMULITE¿ 1000 Third Generation PSA (prostate-specific antigen) kit. The investigation by Siemens has confirmed that patient samples exhibit an average negative bias of approximately -22% with a range of -16% to -27% for samples recovering from 0.08 to 16.2 ng/mL.
Identifiers
| code_info | Lot Numbers: 1) 0336 - Expiration date 2017-01-31 2) 0337 - Expiration date 2017-02-28 3) 0338 - Expiration date 2017-05-31 |
|---|