FDA Device recall Z-1118-2025
UNiD" Adaptive Spine Intelligence UNiD Spine Analyzer Product Number SW3002 UDI-DI code 03613720286929 Two cloud-based software application…
- FDA Device
- Class II
- ongoing
- Published 2025-02-12
On 2025-02-12 the FDA classified a Class II recall of UNiD" Adaptive Spine Intelligence UNiD Spine Analyzer Product Number SW3002 UDI-DI code 03613720286929 Two cloud-based software application… by Medicrea International, citing due to software anomalies that may impact on rod planning specifically, certain optional surgical parameters may have had errors that resulted i incorrect calculations displayed o….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1118-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-12-03 |
| Published | 2025-02-12 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Medicrea International |
| Distribution | AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY |
Product
UNiD" Adaptive Spine Intelligence UNiD Spine Analyzer Product Number SW3002 UDI-DI code 03613720286929 Two cloud-based software applications within UNiD Adaptive Spine Intelligence (ASI): the UNiD HUB and the UNiD Spine Analyzer. The UNiD HUB cloud-based software is a healthcare application to receive, transfer, display, store data needed for planning a spine surgery or for post-operative follow-up (patient information, X-ray image and recommendations for planning). The UNiD Spine Analyzer cloud based software is a healthcare application intended for assisting healthcare professionals in viewing and measuring images as well as planning spine surgeries.
Reason
Due to software anomalies that may impact on rod planning specifically, certain optional surgical parameters may have had errors that resulted i incorrect calculations displayed on system. Impacted parameters includes the following: Roussouly Classification, Real Lumbar Lordosis and Real Thoracic Kyphosis, Barrey Ratio, and Lenke Classification.
Identifiers
| code_info | UNiD HUB is not a medical device and thus does not have a CFN or GTIN code. UNiD Spine Analyzer Product Number SW3002…UNiD HUB is not a medical device and thus does not have a CFN or GTIN code. UNiD Spine Analyzer Product Number SW3002 UDI-DI code: 03613720286929 No lot numbers |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1118-2025%22