FDA Device recall Z-1117-2017

VITROS Na+ Slides, (UPN 10758750004812) For in vitro diagnostic use only

On 2017-02-08 the FDA classified a Class II recall of VITROS Na+ Slides, (UPN 10758750004812) For in vitro diagnostic use only by Ortho Clinical Diagnostics, citing there is a potential for biased results to be generated over the 10 day on-analyzer storage limit when using VITROS Na+ Slide cartridges warmed at room temperature and loaded onto….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1117-2017
ClassificationClass II
Statusterminated
Initiated2016-12-20
Published2017-02-08
Last updated2026-09-13 11:54 UTC
CompanyOrtho Clinical Diagnostics
DistributionUS

Product

VITROS Na+ Slides, (UPN 10758750004812) For in vitro diagnostic use only. VITROS Na+ Slides quantitatively measure sodium (Na+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Reason

There is a potential for biased results to be generated over the 10 day on-analyzer storage limit when using VITROS Na+ Slide cartridges warmed at room temperature and loaded onto the analyzer after using the minimum warm up requirements stated in the IFU. If Quality Control (QC) results are within acceptable limits, reported results are acceptable and were not affected by this issue.

Identifiers

code_infoGeneration (GEN): 1 through 6, 30 through 40, 44 through 49 Expiry Dates: 01-JAN-2017 through 01-JUN-2018

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.