FDA Device recall Z-1112-2025

Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems

On 2025-02-12 the FDA classified a Class II recall of Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems by Beckman Coulter, citing identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of result….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1112-2025
ClassificationClass II
Statusongoing
Initiated2025-01-02
Published2025-02-12
Last updated2026-09-13 11:57 UTC
CompanyBeckman Coulter
DistributionUS

Product

Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems

Reason

Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time required to achieve a passing calibration curve.

Identifiers

code_infoUDI-DI 15099590736002 Lots 339062, 439811

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.