FDA Device recall Z-1110-2019
AquaBeam Handpiece, Sterile EO, Rx Only, REF: 310301, and 320301, a component of AquaBeam Robotic System. 320301 distributed outside the U.…
- FDA Device
- Class II
- terminated
- Published 2019-04-17
On 2019-04-17 the FDA classified a Class II recall of AquaBeam Handpiece, Sterile EO, Rx Only, REF: 310301, and 320301, a component of AquaBeam Robotic System. 320301 distributed outside the U.… by Procept Biorobotics, citing certain lots may develop an internal leak due to a micro-crack in a specific weld joint within the device. When the Robot detects this failure, it immediately renders the device i….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1110-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-03-06 |
| Published | 2019-04-17 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Procept Biorobotics |
| Distribution | US |
Product
AquaBeam Handpiece, Sterile EO, Rx Only, REF: 310301, and 320301, a component of AquaBeam Robotic System. 320301 distributed outside the U.S. Device Description The AQUABEAM Robotic System is intended for use in patients suffering from lower urinary tract symptoms resulting from benign prostatic hyperplasia (BPH).
Reason
Certain lots may develop an internal leak due to a micro-crack in a specific weld joint within the device. When the Robot detects this failure, it immediately renders the device inoperable and notifies the user via an error message, which may cause delay and/or inability to complete the procedure.
Identifiers
| code_info | Domestic Lots: 18C00681, 18C00770, 18C00852, 18C01244. Foreign Lots: 18C00799, 18C00844, 18C00985, 18C01024, 18C01044, 18C01234 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1110-2019%22