FDA Device recall Z-1108-2020
Terumo Advanced Perfusion System 1 Roller Pump, 4 Inch, Catalog Number 816570 - Product Usage: is indicated for use for up to 6 hours in th…
- FDA Device
- Class II
- terminated
- Published 2020-02-19
On 2020-02-19 the FDA classified a Class II recall of Terumo Advanced Perfusion System 1 Roller Pump, 4 Inch, Catalog Number 816570 - Product Usage: is indicated for use for up to 6 hours in th… by Terumo Cardiovascular Systems, citing these roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypica….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1108-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-01-10 |
| Published | 2020-02-19 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Terumo Cardiovascular Systems |
| Distribution | US |
Product
Terumo Advanced Perfusion System 1 Roller Pump, 4 Inch, Catalog Number 816570 - Product Usage: is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical proffessional who is experienced in the operation of this or similar equipment. The roller pump is used to move fluids through the cardiopulmonary bypass circuit and can be used for a number of applicatinos including arterial blood pumping, cardioplegia delivery, suction, and venting.
Reason
These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.
Identifiers
| code_info | UDI: 00886799000533; Serial Numbers: 6860 6861 6862 6863 6864 6865 6876 6877 6885 6910 6911 6912 6913 6934…UDI: 00886799000533; Serial Numbers: 6860 6861 6862 6863 6864 6865 6876 6877 6885 6910 6911 6912 6913 6934 6935 6936 6937 6938 6939 6940 6941 6942 6943 6944 6945 6946 6947 6948 6949 6950 6951 6952 6953 |
|---|