FDA Device recall Z-1107-2017
Genetic Screening Processor (GSP)
- FDA Device
- Class II
- terminated
- Published 2017-02-08
On 2017-02-08 the FDA classified a Class II recall of Genetic Screening Processor (GSP) by Perkinelmer Life And Analytical Sciences Wallac Oy, citing the firm became aware that the ball bushing axels are defective in the GSP instrument. During exposure to humid conditions, there is a risk that the axle material will corrode. De….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1107-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-01-05 |
| Published | 2017-02-08 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Perkinelmer Life And Analytical Sciences Wallac Oy |
| Distribution | AL, MA, SC, TX |
Product
Genetic Screening Processor (GSP); PerkinElmer Part Number= 2021-0010 Intended for in vitro quantitative and qualitative determination of analytes in body fluids.
Reason
The firm became aware that the ball bushing axels are defective in the GSP instrument. During exposure to humid conditions, there is a risk that the axle material will corrode. Deterioration of the axle can be observed as deposits of corrosion on the defective axles. Ball bushing axles may corrode and corrosion particle may cause contamination in the wells or the corrosion will cause movement issues and possibly spillage and causing contamination of the plate wells.
Identifiers
| code_info | Serial Numbers: 20210239 20210243 20210244 20210245 20210246 20210247 20210248 20210249 20210250 20210251 202…Serial Numbers: 20210239 20210243 20210244 20210245 20210246 20210247 20210248 20210249 20210250 20210251 20210252 20210253 20210254 20210255 20210256 20210257 20210258 20210259 20210260 20210261 20210262 20210263 20210264 20210265 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1107-2017%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.