FDA Device recall Z-1106-2026
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 14.0.0, 15.0.0, 15.1.3 Software Version: RayStation 2023B, Ra…
- FDA Device
- Class II
- ongoing
- Published 2026-01-28
On 2026-01-28 the FDA classified a Class II recall of Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 14.0.0, 15.0.0, 15.1.3 Software Version: RayStation 2023B, Ra… by Raysearch Laboratories Ab, citing potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no cont….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1106-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-11-28 |
| Published | 2026-01-28 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Raysearch Laboratories Ab |
| Distribution | n/a |
Product
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 14.0.0, 15.0.0, 15.1.3 Software Version: RayStation 2023B, RayStation 2024A, RayStation 2024A SP3 Product Description: Radiation Therapy Treatment Planning System Component: No
Reason
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
Identifiers
| code_info | Lot Code: UDI: 0735000201055620230630, 0735000201072320231213, 0735000201088420250129 GTIN: 07350002010556, 07350002010…Lot Code: UDI: 0735000201055620230630, 0735000201072320231213, 0735000201088420250129 GTIN: 07350002010556, 07350002010723, 07350002010884 Serial Numbers: 14.0.0.3338, 15.0.0.430, 15.1.3.10 Software Revisions: RayStation 2023B, RayStation 2024A, RayStation 2024A SP3 Expiration Date: 2030-02-21 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1106-2026%22