RecallRadar

FDA Device recall Z-1105-2021

Bemer-Set/ Pro-Set for the following models: REF. 410201 (BEMER Pro-Set), REF. 420201 (B.BOX Professional), REF. 420100 (B.BOX CLASSIC), RE…

On 2021-02-17 the FDA classified a Class II recall of Bemer-Set/ Pro-Set for the following models: REF. 410201 (BEMER Pro-Set), REF. 420201 (B.BOX Professional), REF. 420100 (B.BOX CLASSIC), RE… by Bemer Int Ag, citing there potential that electromagnetic fields generated in the therapy system may cause technical defects in a patient using an insulin pumps or a malfunction of a pacemaker.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1105-2021
ClassificationClass II
Statusterminated
Initiated2021-01-13
Published2021-02-17
Last updated2026-09-13 11:56 UTC
CompanyBemer Int Ag
Distributionn/a

Product

Bemer-Set/ Pro-Set for the following models: REF. 410201 (BEMER Pro-Set), REF. 420201 (B.BOX Professional), REF. 420100 (B.BOX CLASSIC), REF. 410100 (BEMER Classic-Set), REF. 420200 (B.BOX Professional), and REF. 410200 (BEMER Pro - Set)

Reason

There potential that electromagnetic fields generated in the therapy system may cause technical defects in a patient using an insulin pumps or a malfunction of a pacemaker.

Identifiers

code_infoAll Lot Numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.